Product search:

Back

1. Impurity A; 2. Clindamycin phosphate; 3. Impurity E

Separation of clindamycin (QC)

Article number: VPH0006J

Bookmark product

BlueOrchid 175-1.8 C18 A, 100 x 2 mm ID

Order No. 10BI184BOE

Vertex Plus column, 30x2 mm, 50x2 mm

For a link to the column see below: Related products


Details
Method

UHPLC

Eluent

Buffer (13.6 g/lKH2PO4) /

Acetonitrile 80:20 (v/v)

 

Gradient

isocratic

Flow rate

0.7 ml/min

Mode

RP-Mode

Injection volume

5 µl

Pressure

855 bar (12400 psi)

Temperature

30°C

Detection

PDA-1, 210 nm (50 Hz)

10 mm, 2 µ l flow cell

 

Chromatogram

1. Impurity A

2. Clindamycin phosphate

3. Impurity E

 

Downloads
  Title Description Languages
Separation of clindamycin phosphate and process impurities Antibiotics, process impurities, quality control English
Bestimmung von Verunreinigungen im Wirkstoff Clindamycinphosphat Antibiotika, Prozessverunreinigungen, Qualitätskontrolle Deutsch
VPH0006J short application Separation of Clindamycin (QC)
Related products
  Title Article number Description
Vertex Plus Column 100 x 2 mm BlueOrchid 175-1.8 C18A 10BI184BOE Vertex Plus Column 100 x 2 mm BlueOrchid 175-1.8 C18A

Technical data are subject to change without notice.